: A Georgia LawyerSymbiq One- and Two-Channel Infusers Recalled
The manufacturer mailed an updated clinical bulletin on June 11, 2010, which stated that the user does not have to remove or stop using the Symbiq infusion pump, and it provided recommended mitigation actions such as not running solution/medication containers dry and programming the pump for 95% of stated volume of container (minus the priming volume) or less.
To date, no reports of serious injuries have been received, but the manufacturer states that "failure to detect air in line may result in the delivery of air to the patient, resulting in serious injury or death.